SPECIFICATION:
Non-invasive Blood Pressure
Simulation waveform : Oscillometric
Integrated Pump : 0 to 350 mmHg user configurable
Leak test : user configurable between 0-350mmHg
Chronometer : configurable upto 999 secs
Digital Manometer : 0 – 410 mmHg
Pressure Accuracy : +/- 0.5% FS
Pressure Units : mmHg,inHg,kg/cm2,cmH2O,mBar,PSI,inH2O and kPa
Oxygen Saturation
Range : 50 to 100%
Accuracy : ± 1 bpm
Chronometer : test option to test response time SpO2 monitor.
ECG Arrhythmia Simulator
ECG : full 12 lead simulation including high level output
Wave forms : Normal Sinus Rhythm (NSR), ST Elevation, ST
Depression, Myocardial Infarction,Tall T
Rate : 20 – 300 BPM typed value
Performance waveforms : Sine, Square, Triangle, and Pulse
Pacer Waveforms : Synchronous Atrial, Asynchronous Atrial
Pacer Only : Pacer Pulse rate 60 BPM, Ventricular Pacer rate 70 BPM, Atrial &
Ventricular Pacer Heart rate 70 BPM, R Wave Detection
Arrhythmia Waveforms
Atrial : Sinus Arrhythmia (SA), Missing beat, Atrial Flutter (AFLT), Atrial
Fibrillation (AFB), Paroxysmal Atrial Tachycardia (PAT), Junctional
Premature Contraction,
Atrial Conduction : First Degree AV Block, Second Degree AV Block – Mobitz I, Second
Degree AV Block – Mobitz II, Third Degree AV Block, Right Bundle
Branch Block (RBB), Left Bundle Branch, Block, (LBB), Left Anterior
Hemiblock
Ventricular : Premature Ventricular Contraction – Intermittent Premature,
Ventricular Contraction – Continuous, Bigeminy,Trigeminy, Ventricular
Flutter (VFLT),Ventricular Fibrillation (VFB), Ventricular,Tachycardia
(VTC), Right Focal PVC.
Temperature
Simulation : YSI 400 / 700 Static
Range preset at : 25, 33, 37and 41°C
Respiration
Rates : 5, 10, 15, 30, 60, 120, 180 Breaths per Minute
Base resistances : 250, 500, 750, 1000 ohms
Apnoea Simulation : 0 – 60 seconds duration 0 – 300 seconds interval.
Invasive Blood Pressure
Channels : 2 channels
Static : 0 to 300mmHg. Typed values
Dynamic : 0 – 300mmHg for systolic & diastolic General
Operation : Battery cell, insitu charge
Battery charger : 110-230VAC, 50/60 Hz
Battery life : 4-24 hours testing depending on simulations
Memory Capacity : Appr. 5,000 records
Communication : Via Blue Tooth
Display : Monochorme, ¼ VGA full graphics
Keypad : Alpha-numeric
Weight : <1.5 kg
Size : (L x W x D) 270 x 110 x 75 mm / 10.5 x 4 x 3”
Operating conditions : 10-30’C, 0-90% RH – NC
Storage environment : -15’ – +60’C
Environmental Protection : IP 40
Part number : 370A930
Free Calibration certificate Uncertainty Certificate
Standard Accessories : PULS-R (Universal SPO2 Simulation Finger)
Technical Specifications
Accuracy of simulation when used with the corresponding R-curves
Resolution : 1% steps
: 1% steps
: 1% steps
Range : 30-59%
: 60-89%
: 90-100%
Repeatability* : ±5%**
: ±3%
: ±1%
*Based on using the same probe and monitor setup
**Note that some monitor types might not be able to display low range sats
Heart Rate Setting : 30-300 bpm
*** Operating temperature : 0-40°C
Dimensions : 78 x 68 x 29 mm
Finger shape : 40 x 18 mm
Part number 399A910
Standard Accessories : Carrying case,5-pc Tubing Adaptor (NIBP), ECG interface box, Power Supply, 30 day Med-eBase trial, Aplication Software.