SPECIFICATION: | |
Earth Continuity | |
Pre-pulse | 65-25A peak current, (0.1 to 0.8 Ω respectively) |
Pulse shape | Exponential decay |
Decay time | 200 – 550ìs to 5% of peak current, (0.1 to 0.8 Ω respectively) |
Method 2 wire | 2 wire |
Measurement Current | >± 200mADC into 2Ω |
Max Test Voltage | 4-24Vrms o/c |
Measuring Range (low range) | 0.001 – 0.999Ω |
Resolution | 0.001Ω |
Measuring Range (mid range) | 1.00 – 9.99Ω |
Resolution | 0.01Ω |
Measuring Range (high range) | 10.0 – 19.9Ω |
Resolution 0.1Ω | 0.1Ω |
Accuracy ± 1% of value, ± 5mΩ | ± 1% of value, ± 5mΩ |
Circuit Protection | Test inhibited if ≥ 30VAC or DC at 4mm inputs |
Insulation Resistance | |
Measurement | EUT to Earth |
Voltage 500VDC @1mA | 500VDC @1mA |
Maximum O/C Voltage | <600VDC |
Range | 100K – 20MΩ ± 5% ± 2 digits |
Resolution 0.01MÙ | 0.01MΩ |
Short circuit current <2mA | <2mA |
Powered Leakage Measurements | |
IEC 62353 | Equipment Leakage (Direct) |
IEC / AAMI 60601 | Earth + Enclosure Leakage |
NFPA-99 | Ground + Chassis Leakage |
IEC 61010 | Touch Leakage, Touch Voltage |
Test Voltage | Mains Supply Voltage |
Measuring Range | 0.1- 9999μA (0.1-8000μA typical for IEC61010) |
Measurement/Display Resolution | 0.1μA |
Accuracy | ± 2%, ± 5μA |
Mains Reversal | Soft key |
Single Fault Conditions | Open neutral and Open earth via soft key |
Frequency Response | IEC 60601 – 62353, NFPA-99, and IEC 61010 selectable |
Voltage measurement | |
Application | L-N, L-E, N-E and touch voltage (IEC 61010) |
Range | 0.0V – 300VAC |
Resolution | 0.1V |
Accuracy | ± 2% ± 2 digits (between 10V – 270VAC) |
Mains frequency | 45.0 – 66.0Hz |
Resolution 0.1Hz | 0.1Hz |
Accuracy | Unspecified |
EUT Load Current Measurement | |
Range 0.0A – 20.0A | 0.0A – 20.0A |
Resolution 0.1A | 0.1A |
Accuracy ±5% ± 2 digits | ±5% ± 2 digits |
General | |
Mains power | 90-264V 48-64Hz |
Maximum current rating | 20A @ 120V / 16A @ 230V |
Duty cycle (@21°C ambient) | 16A to 20A, 3 min. on/ 10 min. off |
10A to 15A, 3 min. on/ 5 min. off | |
0A to 10A, continuous | |
Weight | 1.1kg / 2.5lbs (unit) |
2.2kg / 5lbs (complete with accessories) | |
Size (L x W x D) | 225 x 150 x 100mm / 9 x 6 x 4” |
Operating temperature | 0 to 40°C |
Humidity | 0 -98% Relative humidity, non-condensating |
Storage temperature | -10 to 50°C |
Altitude 0 – 2000m | 0 – 2000m |
Ingress Protection | IP 40 |
Operating pollution | degree 2, according to IEC 60529 |
Standard Accessories : | |
Detachable mains cable | |
Bond test probe with clip | |
Quick start guide | |
Calibration certificate | |
Carrying case |