Technical Specifications for OxSim OX-1
| Physical Dimensions | |
| Size | 2.5” x 2.5” x 1.25 “ (excluding finger) |
| Weight | 5 ounces (excluding battery) |
| SpO2 Connection | Optical |
| Power | |
| Battery Power | 1 AA battery (any chemistry) |
| Battery Life (Alkaline) | 8 to 10 hours of simulation time (typical, depending on the oximeters under test) 60 days in sleep mode (unit sleeps after 10 min. with no probe connection) |
| AC operation | Included 6V DC supply operates over 100-240 VAC, 50-60 Hz |
| Simulation Accuracy | |
| Saturation Error | Error in the displayed saturation value should be less than or equal to the specified error of the sensor which is typically +/- 2%, but which can be +/-3% or more for some disposable sensors. |
| Heart Rate Error | Less than 1bpm |
| Modes of Operation | |
| 85% | |
| Saturation | 85% |
| Pulse Rate | 80 bpm |
| Perfusion Index | 2.0 |
| 95% | |
| Saturation | 95% |
| Pulse Rate | 40 bpm |
| Perfusion Index | 2.0 |
| 98% | |
| Saturation | 98% |
| Pulse Rate | 80 bpm |
| Perfusion Index | 2.0 |
| 140 BPM | |
| Saturation | 98% |
| Pulse Rate | 140 bpm |
| Perfusion Index | 2.0 |
| Low Perfusion | |
| Saturation | 99% |
| Pulse Rate | 80 bpm |
| Perfusion Index | 0.2 |
| Sensor Type Selection | |
| #1 Auto | Used for:
All Nellcor Masimo HP/Philips (blue extension cable) HP/Philips (tan soft sensor) GE/Marquette Dräger/Siemens Datascope Datex-Ohmeda (SpO2 SW version dated 1990-2001) Protocol Spacelabs Nihon Kohden Nonin Colin Welch Allyn |
| #2 HP/Ohm | Used for: HP cocoa brown soft sensor, older Ohmeda and older Novametrix |
| #3 Other | Datex Ohmeda (SpO2 SW version dated 2001-2005) |
| Environmental | |
| Operating | Temperature: 10°C to 35°C (50°F to 95°F), Relative Humidity: 10 to 80% non condensing, Altitude: 3,000m (9,843ft) |
| Storage | -20°C to +60°C (-4°F to +140°F) |