SPECIFICATION: 

Earth Continuity

Method 2 wire technique, using ‘zero’ lead function. 

Test Current >+200mA, -200mA DC into 2 ohms 

Max Test Voltage 4-24V rms o/c (6V for IEC 60601 tests) 

Measuring Range (low range) 0.001 – 0.999 ohms @ 0.001 ohms resolution 

Measuring Range (mid range) 1.00 – 9.99 ohms /@ 0.01 ohms resolution 

Measuring Range (high range) 10.0 – 19.9 ohms @ 0.1 ohms resolution 

Accuracy ± 3% of reading + 10 m ohms 

 

Insulation Resistance

Measurement EUT to Earth / Ground, EUT to AP, AP to Ground 

Voltage 250V DC, 500 V DC @1mA. 

Range (low range) 0.01Mohms – 20 Mohms 

Accuracy (low range) ± 5% of reading +2 counts 

Range (high range) 20Mohms – 100Mohms 

Accuracy (high range) ±10% +2 counts 

Resolution 0.01Mohms 

 

Direct Leakage Measurement

Measuring Range 4µA to 9999µA 

Accuracy ± 5% or reading +2 counts 

Mains on A.P. voltage F-type only @ 110% of mains 

Measuring Device As per IEC 60601-1 requirements 

Measurement Type Separate AC & DC for Patient (-Auxiliary) True RMS for all remaining 

Leakage tests Leakage to IEC 60601 

 

Differential Leakage Measurement

Measuring Range 75µA to 9999µA 

Accuracy ±5% of reading + 5 counts 

Measurement / display resolu 1µA 

Measurement Type True RMS 

Measuring Device Similar frequency response characteristics to IEC 60601-1. 

 

Alternative Leakage Measure 

Test Voltage 250V at mains frequency 

Test Current 3.5 mA current limited 

Measurement Range 4µA to 9999µA 

Measurement Resolution 1µA 

Measurement Accuracy ±5% of reading + 2 counts 

Measurement Type True RMS 

Measuring Device As per IEC 60601-1 

 

Power Measurement

Method VA rating. 

Range 0.1KVA – 4KVA 

Accuracy ±10% + 2 counts 

 

Mains Outlet Test

Input voltage range 0-300V AC, max current 16A 

Measures L – E, N – E & L – N 

Accuracy ± 5% of reading + 2 counts 

 

IEC Mains Lead Test

Test Duration 2s 

Test: Continuity of all conductors, Earth bond, Insulation & Polarity 

 

General

Mains power 230 VAC ±10%, 50Hz +/- 1Hz 120 VAC ±10%, 60Hz +/- 1Hz (USA model) 

Battery 6 x 1.5V Alkaline AA 

Weights 1.6 kg including batteries 

Size (L x W x D) 270 x 110 x 75 mm / 10.5 x 4 x 3” 

Operating conditions 0′ – 40’C, 0-90% RH – NC 

Storage environment -15′ – +60’C 

Environmental Protection IP 40 

Calibration Certificate 

Uncertainty Certificate 

 

Included standard accessories

Carrying Case

Earth bond test probe with clip

Earth bond clip lead

Patient Applied part module

10 applied part adaptors

Detachable 2 meter mains cable 