SPECIFICATION:
Earth Continuity
Method 2 wire technique, using ‘zero’ lead function.
Test Current >+200mA, -200mA DC into 2 ohms
Max Test Voltage 4-24V rms o/c (6V for IEC 60601 tests)
Measuring Range (low range) 0.001 – 0.999 ohms @ 0.001 ohms resolution
Measuring Range (mid range) 1.00 – 9.99 ohms /@ 0.01 ohms resolution
Measuring Range (high range) 10.0 – 19.9 ohms @ 0.1 ohms resolution
Accuracy ± 3% of reading + 10 m ohms
Insulation Resistance
Measurement EUT to Earth / Ground, EUT to AP, AP to Ground
Voltage 250V DC, 500 V DC @1mA.
Range (low range) 0.01Mohms – 20 Mohms
Accuracy (low range) ± 5% of reading +2 counts
Range (high range) 20Mohms – 100Mohms
Accuracy (high range) ±10% +2 counts
Resolution 0.01Mohms
Direct Leakage Measurement
Measuring Range 4µA to 9999µA
Accuracy ± 5% or reading +2 counts
Mains on A.P. voltage F-type only @ 110% of mains
Measuring Device As per IEC 60601-1 requirements
Measurement Type Separate AC & DC for Patient (-Auxiliary) True RMS for all remaining
Leakage tests Leakage to IEC 60601
Differential Leakage Measurement
Measuring Range 75µA to 9999µA
Accuracy ±5% of reading + 5 counts
Measurement / display resolu 1µA
Measurement Type True RMS
Measuring Device Similar frequency response characteristics to IEC 60601-1.
Alternative Leakage Measure
Test Voltage 250V at mains frequency
Test Current 3.5 mA current limited
Measurement Range 4µA to 9999µA
Measurement Resolution 1µA
Measurement Accuracy ±5% of reading + 2 counts
Measurement Type True RMS
Measuring Device As per IEC 60601-1
Power Measurement
Method VA rating.
Range 0.1KVA – 4KVA
Accuracy ±10% + 2 counts
Mains Outlet Test
Input voltage range 0-300V AC, max current 16A
Measures L – E, N – E & L – N
Accuracy ± 5% of reading + 2 counts
IEC Mains Lead Test
Test Duration 2s
Test: Continuity of all conductors, Earth bond, Insulation & Polarity
General
Mains power 230 VAC ±10%, 50Hz +/- 1Hz 120 VAC ±10%, 60Hz +/- 1Hz (USA model)
Battery 6 x 1.5V Alkaline AA
Weights 1.6 kg including batteries
Size (L x W x D) 270 x 110 x 75 mm / 10.5 x 4 x 3”
Operating conditions 0′ – 40’C, 0-90% RH – NC
Storage environment -15′ – +60’C
Environmental Protection IP 40
Calibration Certificate
Uncertainty Certificate
Included standard accessories :
Carrying Case
Earth bond test probe with clip
Earth bond clip lead
Patient Applied part module
10 applied part adaptors
Detachable 2 meter mains cable