Technical Specifications for OxSim OX-1
Physical Dimensions | |
Size | 2.5” x 2.5” x 1.25 “ (excluding finger) |
Weight | 5 ounces (excluding battery) |
SpO2 Connection | Optical |
Power | |
Battery Power | 1 AA battery (any chemistry) |
Battery Life (Alkaline) | 8 to 10 hours of simulation time (typical, depending on the oximeters under test) 60 days in sleep mode (unit sleeps after 10 min. with no probe connection) |
AC operation | Included 6V DC supply operates over 100-240 VAC, 50-60 Hz |
Simulation Accuracy | |
Saturation Error | Error in the displayed saturation value should be less than or equal to the specified error of the sensor which is typically +/- 2%, but which can be +/-3% or more for some disposable sensors. |
Heart Rate Error | Less than 1bpm |
Modes of Operation | |
85% | |
Saturation | 85% |
Pulse Rate | 80 bpm |
Perfusion Index | 2.0 |
95% | |
Saturation | 95% |
Pulse Rate | 40 bpm |
Perfusion Index | 2.0 |
98% | |
Saturation | 98% |
Pulse Rate | 80 bpm |
Perfusion Index | 2.0 |
140 BPM | |
Saturation | 98% |
Pulse Rate | 140 bpm |
Perfusion Index | 2.0 |
Low Perfusion | |
Saturation | 99% |
Pulse Rate | 80 bpm |
Perfusion Index | 0.2 |
Sensor Type Selection | |
#1 Auto | Used for:
All Nellcor Masimo HP/Philips (blue extension cable) HP/Philips (tan soft sensor) GE/Marquette Dräger/Siemens Datascope Datex-Ohmeda (SpO2 SW version dated 1990-2001) Protocol Spacelabs Nihon Kohden Nonin Colin Welch Allyn |
#2 HP/Ohm | Used for: HP cocoa brown soft sensor, older Ohmeda and older Novametrix |
#3 Other | Datex Ohmeda (SpO2 SW version dated 2001-2005) |
Environmental | |
Operating | Temperature: 10°C to 35°C (50°F to 95°F), Relative Humidity: 10 to 80% non condensing, Altitude: 3,000m (9,843ft) |
Storage | -20°C to +60°C (-4°F to +140°F) |