Technical Specifications for OxSim OX-1

Physical Dimensions
Size  2.5” x 2.5” x 1.25 “ (excluding finger)
Weight  5 ounces (excluding battery)
SpO2 Connection  Optical
Power
Battery Power  1 AA battery (any chemistry)
Battery Life (Alkaline)  8 to 10 hours of simulation time (typical, depending on the oximeters under test) 60 days in sleep mode (unit sleeps after 10 min. with no probe connection)
AC operation  Included 6V DC supply operates over 100-240 VAC, 50-60 Hz
Simulation Accuracy
Saturation Error  Error in the displayed saturation value should be less than or equal to the specified  error of the sensor which is typically +/- 2%, but which can be +/-3% or more for  some disposable sensors.
Heart Rate Error  Less than 1bpm
Modes of Operation
85%
Saturation  85%
Pulse Rate  80 bpm
Perfusion Index  2.0
95%
Saturation  95%
Pulse Rate  40 bpm
Perfusion Index  2.0
98%
Saturation  98%
Pulse Rate  80 bpm
Perfusion Index  2.0
140 BPM
Saturation  98%
Pulse Rate  140 bpm
Perfusion Index  2.0
Low Perfusion
Saturation  99%
Pulse Rate  80 bpm
Perfusion Index  0.2
Sensor Type Selection
#1 Auto  Used for:  

All Nellcor 

Masimo 

HP/Philips (blue extension cable) 

HP/Philips (tan soft sensor) 

GE/Marquette 

Dräger/Siemens 

Datascope 

Datex-Ohmeda (SpO2 SW version dated 1990-2001) 

Protocol 

Spacelabs 

Nihon Kohden 

Nonin 

Colin 

Welch Allyn

#2 HP/Ohm  Used for: HP cocoa brown soft sensor, older Ohmeda and older Novametrix
#3 Other  Datex Ohmeda (SpO2 SW version dated 2001-2005)

 

Environmental
Operating  Temperature: 10°C to 35°C (50°F to 95°F), Relative Humidity: 10 to 80% non condensing, Altitude: 3,000m (9,843ft)
Storage  -20°C to +60°C (-4°F to +140°F)